5
Service Menu Item
Installation Item
Thermo Scientific
TriPlus RSH User Guide
211
The following example describes calibration of the SPME tool (SPME) positioned in Slot 1,
the Headspace Tool (HS) positioned in Slot 2, and the Liquid Tool (LS) positioned in Slot 3.
To calibrate the tools
1. Select the wizard item
Calibrate Tools
and follow the dialog. See
Figure 153.
Wizard Item Calibrate Tools and Calibration of Ideal Tool
2. Press
Save
to continue in the wizard. The X-, Y-, and Z-positions will be stored.
Select
Next
to open the window for assigning the Tool parameters.
The default value of the Tool Parameter is
Not Specified
. This means that the parameter
has to be specified for the application. The definition can be made at this stage of
installation or later on during routine work. You are able to change this parameter to
adapt it for another application; note that the
Extended User Level
is required.
Table 93.
Calibrate Tools Parameters
Item
Description
Not Specified
Prohibits the execution of a run.
Magn10or20mL
Magnetic Ring installed to enable the transport of 10 or
20 mL vials.
Magn2mL
Magnetic Ring not installed but Lower Needle Guide has
Magnets installed to enable transport of a 2 mL vial.
Foil Cutter
Foil Cutter installed. The Foil Cutter pierces a sealing foil of a
MTP or DW-Plate before the syringe needle penetrates the
well.
NonMagn2mL
The lower Needle Guide has no magnets installed; a magnetic
ring is not connected to it. Typically used for Microtiter Plates
or for vials with a non-magnetic cap.