Setup Menu Item
Setup Menu Item Modules
172
TriPlus RSH User Guide
Thermo Scientific
Penetration Speed
The syringe needle penetration speed is set by this parameter.
It is recommended to use the default value first. Changing
this parameter in one or the other direction might lead to
needle bending.
The default values are:
• 40 mm/s in Wash position
• 40 mm/s in Waste position
Z-Tolerance
A tolerance window to give a plus/minus range (expressed in
mm) where the Head must expect an object.
If the value of this item is
> 0
, this indicates a relative
detection mode, and the sensor from the Head needle guide
is active.
If the value of this item is set to
0
the Needle Guide Motor
detection current is turned off, and the syringe slider moves
to an absolute value, Z-axis position as specified in the
object.
Diameter
Specific for Waste Vial. Defines the diameter of the reservoir
container. The default value is 20 mm
Height
Specific for Waste Vial. Defines the height of the reservoir
container. The default value is 80 mm
Vial Type
Defines the Vial Type of the reservoir container in Positions
or the corresponding Waste Type in Waste Position.
Selection for Wash Positions are: Vial Types None, 2SR,
10SR, or 100SR.
Selection for Waste Position are:
•
None
—
If None is selected, an error (Pending Message)
will be propagated.
•
Waste Port
— Large Wash and Fast Wash Module
•
Waste Vial
— Standard Wash.
•
Waste Inset
— Standard Wash with inserted Drainage
Adapter.
Volume
Defines the volume of the reservoir vial.
Table 81.
Parameters for Fast Wash Stations (Sheet 3 of 3)
Item
Description