Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
171
Max Flow Rate
Sets the upper limit of flow rate for the selected
Wash
Position.
Post Rinse Liner
Activates the solvent pump to rinse the wash liner of the
selected
Wash
Position. Number of rinse cycles is selectable
after the syringe needle has been removed from the Wash
Station.
Note
Without the syringe needle inserted in the liner, the
solvent flow is unrestricted. Expect a higher solvent
consumption.
Pre Rinse Liner
Activates the solvent pump to rinse the wash liner of the
selected
Wash
Position. Number of rinse cycles is selectable
before the syringe needle has been inserted into the Wash
Station.
Note
Without the syringe needle inserted in the liner, the
solvent flow is unrestricted. Expect a higher solvent
consumption.
Def Ndl Penetr Depth
Sets the default value of the syringe needle penetration depth
in the reservoir vial.
The default values are:
• 25 mm in Wash position
• 20 mm in Waste position
Depenetration Speed
Selects the required speed for pulling the syringe needle out
of the reservoir vial.
The default values are:
• 100 mm/s in Wash position
• 100 mm/s in Waste position
Max Penetration Depth
The maximum allowed Needle Penetration can be defined as
a safety measure so that the you cannot destroy the needle tip
by going too far down.
The maximum needle penetration must not exceed the
length of the syringe needle, considering the loss of
approximately 12 mm due to needle guide, and so on.
For a typical needle length of 57 mm or 85 mm for the long
needle, the maximum penetration depth must not exceed
45 mm or 73 mm.
Table 81.
Parameters for Fast Wash Stations (Sheet 2 of 3)
Item
Description