3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
139
Trap Heater
This menu list item is specific for the ITEX Tool. Selecting
the item opens the parameters for:
•
Actual Temperature
— Reads out the actual
temperature of the trap; read-only function.
•
Max Temperature
— Sets the upper temperature limit.
The Maximum Temperature which can be set is 350 °C.
•
Min Temperature
— Sets the lower temperature limit.
The Minimum Temperature which can be set is 100 °C.
•
Standby Temperature
— The Standby Temperature
can be adjusted to the application’s requirements.
Setting the Standby Temperature using this function
does not yet activate the newly-selected temperature.
To activate it, go to the pull-up menu
Options
and
select
Set Standby Temperature
.
The Standby Temperature is activated when the TriPlus RSH is
powered up. In routine use, the Standby Temperature is activated
after a job has been terminated. The Standby Temperature is
typically lower than the application temperature.
•
Temp Setpoint
— Read out the setpoint of the
temperature.
•
Stdby Temp Control
—
If switched off, the module will
not go to standby temperature if idle.
•
Temp Gain
— The coefficient used, if a temperature
calibration has been performed. Possible values are
between 0 and 2. Read-only.
•
Temp. Offset
— The Offset temperature used, if a
temperature calibration has been performed. Possible
values are between -10 °C and +10 °C. Read-only.
Location
Location (Slot Position) where the specific tool is parked.
Length
Defines total length of the tool. Read-only function.
Instrument Length
Defines total length of the syringe (instrument). Read-only
function.
Needle Guide Type
This parameter defines if an adapter, such as the magnetic
ring or the foil cutter, is connected to the Needle Guide.
Important:
The Needle Guide Type has to be defined,
default value is
Not Specified
.
Table 66.
List Item Parameters for HS, LS, SPME, and ITEX Tools (Sheet 3 of 4)
Item
Description
Summary of Contents for TriPlus RSH