3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
133
Bottom Sense Parameter: Criteria for Syringe Needle
In order to use the parameter
Bottom Sense
successfully, consider the following points:
• For successful bottom detection, the syringe needle has to be intact (not bent from
previous use).
• The needle length does not exceed 65 mm. If longer needles are used, for example 80 mm
in length, then the needle gauge 26 and 26S (outer needle diameter 0.47 mm) is too thin
to withstand the touchdown pressure.
The needle gauge has to be changed to 23 or 23S (outer needle diameter 0.63 mm).
Syringe Types Options Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
.
Point Style
Specify the syringe suppliers specific.
Scale Length
Specify the length of the scale in mm
Syringe Volume
Specifies the volume of the syringe in
μ
L.
Gas Inlet Pos
Specifies the position in mm of the gas inlet. HS Tool only.
Fiber Length
Specifies the length in mm of the SPME fiber.
Fiber Type
Specifies the type of the SPME fiber. SPME Tool only
Table 62.
List of Syringe Types Parameters (Sheet 2 of 2)
Item
Description
Table 63.
Option Pull-up Menu for Syringe Types
Item
Description
New Syringe Type
Creates a new Syringe Type template.
Copy Syringe Type
Copies the present Syringe Type template.
Export Syringe Type
Exports the present Syringe Type template to a USB stick.
Export All
Exports all Syringe Type templates at once to a USB stick.
Delete Syringe Type
Deletes the present Syringe Type template.