List Items Agitator Reserved Positions and Vial Type
With the list item
Setup Module
it is possible to reserve one or more positions in the Agitator
for a source vial (for example to provide an internal standard solution, a reagent solution for a
derivatization, or any other solution) as required by a specific application. Possible reserved
positions for source vials begin at
Vial Position 6
and count backwards.
The vial size can be selected for the remaining free positions and can be selected separately for
the reserved vial(-s).
Note that only one vial size can be selected for the reserved vials. The consequence is that
during routine applications the vial overlapping will be possible only within the free available
sample vial positions to optimize the throughput. The behavior of the following two
Activities
is summarized:
•
Transport Vial
activity — Allowed for available free sample vial positions and for
reserved vial positions.
•
Move To Object
activity — Allowed for reserved vial positions only.
The reason for this limitation is because in the overlapping mode the sample vial position is
managed by the scheduler.
You can change the
Vial Type
at any time during a routine application selecting the menu
item
Agitator
, scrolling down to the list item
Vial Type
and selecting another Vial Type.
Examples of Free available and Reserved Vial Positions are visualized in the following images: