TriPlus RSH Basic Informations
Starting the TriPlus RSH
84
TriPlus RSH User Guide
Thermo Scientific
When an error is observed, check all connections, try to eliminate the source of the
problem, and restart the TriPlus RSH again.
The messages accompanied by the
Exclamation Mark
warning symbol indicate that a
module has not been completely configured.
shows an example where the Vial
Type has not been set for the Solvent Reservoir, or for Wash Vial#1.
See also the information provided by the error message or consult the chapter
Troubleshooting
in the
TriPlus RSH Hardware Manual
.
Short-cuts from Startup Screen
The following steps are required to use the short-cuts:
1. Position the cursor on the list item at the start screen for which you wish to check certain
parameters.
2. Press the
Enter
button to open the dialog for the selected item.
3. Selecting the
Options
pull-up menu, it provides a list of specific tasks, limited to the
typical needs for routine work.
The example in
shows the resulting errors when the Vortexer cable is not
connected, and the homing could not be performed. All the subsequent boot-up checks
failed.
Figure 61.
List Items of Pending Messages
Note
Selecting a menu list item from the startup screen provides direct access, and offers
the same functionality as when selecting the item using the path
Menu Button
|Options |
Setup | RSH Modules
.
Note
Parameters which can be modified are described in detail in
. Parameters which have read-only access are often self-explanatory and are usually
only mentioned in the listing.