Setup Menu Item
Setup Menu Item Modules
110
TriPlus RSH User Guide
Thermo Scientific
Injectors Options Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
.
OC Actuator
The OC Actuator allows the opening and closing of the On-column Injector rotary valve
when the syringe needle is inserted or removed.
The OC Actuator is an active module, which means that the
OC Electronic Box
is connected
to the Bus and the cable from the GC Injector is connected to the Box. For details see the
section
.
Selecting the list item
OC Actuator
opens up the following parameters. See
.
Z Tolerance
A tolerance window to give a plus/minus range (expressed in
mm) in which the Head must expect an object.
• If the value of this item is
> 0
, indicates a relative
detection mode is triggered, and the sensor from the
Head needle guide is active.
• If the value of this item is set to
0
, the sensor of the
needle guide is turned off and the syringe slider moves to
an absolute value: the Z-axis position as specified for the
object.
Table 41.
Injector Parameters (Sheet 2 of 2)
Item
Description
Table 42.
Option Menu for Injectors
Item
Description
Reset to Default
Resets the selected module parameter to default value.
Teach Module
The positions of the X-, Y-, and Z-axes can be defined again,
(teaching).
Check Teaching
Checks the teaching position of the module. This task serves
as a quick check to determine if the X-, Y-, and Z-positions
are defined correctly. In case a teaching point is incorrect,
return to the Menu
Setup
for readjustment of the teaching
position.
This action can be done under the
Extended User Level
but
not under
User Level
.
Note
The OC Adapter is part of the GC On-column Injector and it is not provided with
the TriPlus RSH.