3
Setup Menu Item
Setup Menu Item Modules
140
TriPlus RSH User Guide
Thermo Scientific
Tools Options Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
.
Syringe Type
Use the ITEX syringe only.
Parameters for LS and SPME Tools
Location
Location (Slot Position) where the specific tool is parked.
Instrument Length
Defines total length of the syringe (instrument). Read-only
function.
Length
Defines total length of the tool. Read-only function.
Needle Guide Type
Defines if an adapter, such as the magnetic ring or the foil
cutter, is connected to the Needle Guide.
Important:
The Needle Guide Type has to be defined,
default value is
Not Specified
.
Needle Penetration Depth
Defines the penetration depth of the fiber in the vial.
Syringe Type
All syringes with the same geometry (barrel diameter, needle
length) and corresponding syringe type to the tool are
filtered by the software and are selectable for use.
Table 66.
List Item Parameters for HS, LS, SPME, and ITEX Tools (Sheet 4 of 4)
Item
Description
Table 67.
Options Parameters for HS, LS, SPME, and ITEX Tools
Item
Description
Calibrate Tool
Calibrates the specified tool.
Note
This might be necessary after a tool has been
accidently dropped, and assuming that the Needle Guide
was bent. A
re-calibration
corrects for the difference.
Check Tool Calibration
A calibration check is activated.
Select Tool
Selects another tool. The current tool is parked first and the
new required one is picked up.
Purge Syringe
Available only with HS and ITEX Tools. Allows to purge
manually a headspace syringe.
Park Tool
Parks a tool without picking up another tool from the ATC
Station.
Configure SPME Tool
This line is visualized when a SPME tool is selected.
This task allows to configure a SPME tool. Fiber
adjustment.