3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
169
Select the required menu list item to access the parameters. The wash stations differ in
number of positions, Wash Vial size.
• The
Standard Wash Station
accepts 10 mL vials.
• The
Large Volume Wash Station
contains two 100 mL vials, and a Waste position with a
drainage tubing.
• The
Fast Wash Station
contains active pumps to deliver two different wash solvents
The available menu list parameters are identically for the
Standard
and
Large Solvent Wash
Stations
, asides from the number of positions. See
for details.
The menu list parameters for the
Fast Wash Station
are detailed in
Table 80.
Parameters for Standard and Large Volume Wash Stations (Sheet 1 of 2)
Item
Description
Pos 1 to Pos 4
Positions
1
and
2
are available for the
Large Solvent Wash
Station
.
Positions
1
to
4
are available for the
Standard Wash Station
.
Each Position can be defined by using the next parameters.
Waste
Defines the Waste position.
Max Penetration Depth
The maximum allowed Needle Penetration can be defined as
a safety measure so that the you cannot destroy the needle tip
by going too far down.
The maximum needle penetration must not exceed the
length of the syringe needle, considering the loss of
approximately 12 mm due to needle guide, and so on. For a
typical needle length of 57 mm or 85 mm for the long
needle, the maximum penetration depth must not exceed
45 mm or 73 mm.
Def Ndl Penetr Depth
This parameters sets the default value of the syringe needle
penetration depth in the reservoir vial.
Depenetration Speed
This parameter allows you to select the required speed for
pulling the syringe needle out of the reservoir vial.
Penetration Speed
The syringe needle penetration speed is set by this parameter.
It is recommended to use the default value first.
Changing this parameter in one or the other direction might
lead to needle bending.