1
Getting Familiar with Your TriPlus RSH
HS Tool
Thermo Scientific
TriPlus RSH User Guide
25
Automated
Bottom Sensing
with a syringe needle length of 85 mm is not recommended.
High risk of needle bending.
HS Tool
The Headspace Tool (HS Tool) is used for the Headspace.The HS Tool is provided as a single
block, easy to use to install a syringe and the purge gas line. See
Figure 12.
HS Tool
The syringe temperature is controlled from 40 °C up to 150 °C, in 1 °C steps. The syringe is
flushed with inert gas flow through X-Y-Z-axis.
Besides the coupling pivots for the needle guide and the tool, the purge gas connector is
directly installed on the cover plate of the tool.
The syringe flange is supported by the polymer counterpart to the retaining nut. In the
heating cylinder there are two O-rings which seal the gas flow to enable a syringe purge. It is
recommended to replace these O-rings during preventative maintenance.
PKD injector
• PKD injector — 57 mm needle length LS tool
Table 5.
LS Tool - Injector Compatibility (Sheet 2 of 2)
Injector