Agitator Options Pull-up Menu
Selecting the pull-up menu
Options
opens up following tasks. See
.
Table 32.
Agitator Pull-up menu Options
Item
Description
Check Teaching
Checks the teaching position of the module.
This task serves as a quick check to determine if the
X-, Y-, and Z-positions are defined correctly.
In case a teaching point is off, return to the Menu
Setup
for readjustment of the teaching position.
Please note that this action can be done under the
Extended User Level
but not under
User Level.
Setup Module
Allows the selection of the
Cover Plate Type
, to reserve
one or more vials as a source vial (e.g. for a reagent), and
to define the Vial Type for the 6 positions.
See the section
“List Items Agitator Reserved Positions
Teach Module
The positions of the X-, Y-, and Z-axes can be defined
again, teaching.
Activate Standby Temperature
Activates the newly-set Standby Temperature.
Check Agitator
Starts the spinning motion of the Agitator.
It is used to test the Agitator function, or the spinning
speed for particular application.
Settings are:
•
Options
: On; Off; Home
•
Speed
: Select the required speed, range from 250 rpm
to 750 rpm
Press
Run
to start the Agitator
To stop the motion, press
Options
again, select
Check
Agitator
and set the parameter
Options
to
Off
.
Press
Run
.