Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
163
Vial Type Options Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
.
Strip Distance
When a magnetic cap is used, the distance to move the
TriPlus RSH Head to the side (in the Y-direction) to wipe
off the vial can be defined. Measured in mm.
Volume
This parameter defines the volume of the vial.
Z-Tolerance
A tolerance window to give a plus/minus range (expressed in
mm) where the Head must expect an object.
If the value of this item is
> 0
, this indicates a relative
detection mode, and the sensor from the Head needle guide
is active.
If the value of this item is set to
0
, the sensor of the needle
guide is turned off, and the syringe slider moves to an
absolute value, Z-axis position as specified in the object.
IMPORTANT
Using the SPME technique pay special attention to the parameters
DefNdlPenetrDepth
and
MaxNdlPenetrDepth
.
The penetration depth is always the penetration of the needle itself. Consider the length of
the SPME fiber, 10 mm or 20 mm length. When the fiber is lowered, it must not touch
the bottom of the vial.
Table 76.
Vial Types Parameters (Sheet 3 of 3)
Item
Description
Table 77.
Vial Type Options Pull-up Menu
Item
Description
New Vial Type
Creates a new Vial Type.
Copy Vial Type
Copies the present Vial Type.
Delete Vial Type
Deletes a present Vial Type.
Reset to Default
Resets the selected module parameter to the default value.