4
Maintenance Menu Items
Copy Backup
Thermo Scientific
TriPlus RSH User Guide
187
The command is available in all access levels. After successful creation of the Configuration
Backup, copy the file directly onto the USB Stick. The creation of the Configuration Backup
file does take approximately one to two minutes. The file size is in the range of 1.3 MB,
depending on the TriPlus RSH configuration. The file extension is
*.pack
.
The
Configuration Backup
file can also be used for a
Restore
procedure.
Copy Backup
It is good practice to store the configuration backup in a safe place, if possible on a
network-server. You must keep an authorized configuration backup outside of your premises.
To copy a configuration backup from the SD card, insert the USB stick into the connector
USB Host on the back of the TriPlus RSH.
Refer also to the chapter
RSH Firmware
in the
TriPlus RSH Hardware Manual
.
Note
Be aware that only one backup file can be stored on the SD card.
If for example a Diagnostic Backup file is created and directly followed by creating a
Configuration Backup file, the first file (in this example the Diagnostic Backup file) will
be overwritten. To save both files, it is necessary to copy each file immediately onto the
USB Stick