1
Getting Familiar with Your TriPlus RSH
SPME 1 and SPME 2 Tools
Thermo Scientific
TriPlus RSH User Guide
29
Syringes for SPME Tools
SPME syringe fiber Tool includes a combination of four different fibers for starting-up.
Definition of SPME Penetration Depth
Up to
RSH Firmware version 1.5.0
, the definition of the parameter for
SPME Penetration
Depth
has been defined without including the
Fiber length
.
Starting with
RSH Firmware version 1.6.0
the definition has been changed to reflect the
total length, including the Fiber length. See
.
Figure 17.
SPME Penetration Depth Parameter Comparison
This change in the definition is valid for the SPME Penetration Depth for the following
classes:
• Vial
• Injector
• SPME Conditioning Module
As a result this change in definition affects several method parameters for existing methods.
When working with RSH Firmware version 1.6.0, to achieve the same depth in the fiber
exposure as in the previous RSH Firmware version, the following method parameters have to
be changed.
• Class Vial Depth: Parameter
Sampling depth mode
• Class Injection: Parameter
Injector depth
• Class Fiber Conditioning | Derivatization Vial: Parameter
Derivatization depth
.