5
Service Menu Item
Installation Item
Thermo Scientific
TriPlus RSH User Guide
219
To check the teaching modules
1. Select the wizard item
Check Teaching Modules
. See
.
Figure 158.
Wizard Item Check Teaching Modules
2. Press
Check
to activate the wizard. The TriPlus RSH Head moves to the first module in
the list, positions itself at the Teaching Point, and waits for the next command. This gives
the operator the chance to verify the position, and if necessary make some fine tuning and
save the new settings.
3. Modules which have no direct access to the teaching point (lunette) such as the Agitator
or the Vortexer are treated the same way as it would for a routine run.
In other words, the cover of the Agitator is opened automatically or in combination with
the MHE Adapter it moves the TriPlus RSH Head to the side to stripe off the magnetic
adapter.
4. Press
Next
to continue within the wizard; the TriPlus RSH Head moves to the next
module.
5. The pull-up menu
Previous
moves one step back to the previous module and verifies the
teaching position again.
Note
In this section a fine tuning is not foreseen. If a fine tuning is necessary, go either
, or from the start screen activate the
pull-up menu
Option
, and select
Setup | Modules
.
Note
It is mandatory to insert a vial into the Agitator or the Vortexer to teach the X-,Y-,
and Z-Axes positions. If the vial is not inserted, the Z-Axis value could be out of range,
resulting in an error.