Service Menu Item
Installation Item
Thermo Scientific
TriPlus RSH User Guide
191
Installation Item
Press the
Options
button to open the pull-up menu, and select the menu item
Service
; then
select the list item
Installation
to start the wizard. See
Figure 135.
Selecting the Menu List Item Installation
The
Installation
wizard begins with the basic setup items. Some of the list items can be
accessed under the menu item
Setup
.
The
Installation
wizard includes twelve points. See
. When the TriPlus RSH is
powered up the very first time, specific sections are grayed out until the section Calibration
Motors has been completed. Each point ticks off as it is completed.
When the TriPlus RSH is powered on for the very first time, these wizard items which have to
be accomplished are active (black). The grayed out wizard items will be active as soon as the
system is calibrated (Calibrate Motors) and if the hardware configuration of the TriPlus RSH
requires these points.
Note
During a new installation of the TriPlus RSH, it is advisable to first configure the
Passive modules before activating the
Installation Wizard Service
. One wizard item is
Teaching Module. If the Passive modules, such as the Injector, Wash Module, and so on,
are not yet configured, it will be necessary to restart this wizard again.
The
Quick Reference Guide
in the
TriPlus RSH Hardware Manual
shows the order of
installation. Use this guide as reference.