9 • Troubleshooting
Doc. 065250-06 03/2020
111
This error occurs if the driver current is above the allowed value.
To troubleshoot:
1. Turn off the pump flow.
2. Turn off the SP power briefly by pressing the SP power switch.
3. Turn on the SP power again and restart the pump flow.
4. If the problem persists, contact Technical Support for Dionex products for
assistance.
This error occurs if the loading pump electronics fail to detect when the pump
reaches the home position. This can be caused by high pressure in the flow
path.
To troubleshoot:
1. Check for a blockage in the pump outlet tubing or in tubing downstream
from the pump (see
for the SP1 flow schematic; see
for the SP2 flow schematic).
2. If no blockage is found, turn off the pump flow.
3. Turn off the SP power briefly by pressing the SP power switch.
4. Turn on the SP power again and restart the pump flow.
5. If the problem persists, contact Technical Support for Dionex products for
assistance.
Loading pump driver is over current
Loading pump home sensor not detected
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...