Dionex Integral Systems Operator’s Manual
132
Doc. 065250-06 03/2020
10.5 Replacing a Pump Piston
Continued leaking of the piston seal after installation of a new seal (assuming that
the pump head is tight) indicates a dirty, scratched, or broken piston.
Follow the instructions in
to install a new piston (P/N 066110) and
piston seal (P/N 069006) in the dilution pump or the stepper motor loading pump
(P/N 068561, both pumps). When replacing a piston, always replace the piston
seal, also.
10.6 Replacing the Waste Valve Seal
A damaged seal causes leakage around the base of the waste valve knob.
1. Turn off the main power.
2. Disconnect the power cord.
3. To remove the waste valve from the pump head (see
knob counterclockwise until it is loose, and then pull the knob straight out of
the cavity in the pump head.
Figure 10-5. Waste Valve
Waste Valve
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...