7 • Startup, Operation, and Shutdown
Doc. 065250-06 03/2020
91
Calculate the diluted standard concentration as follows:
Where:
Standard loop = V
1
Dilution volume = V
2
Dilution factor = D
f
Calibration standard concentration = C
1
Diluted standard concentration = C
2
13. Install the loop between ports 1 and 6 of the metering (ME) valve.
Analytical Pump Calibration
Calibrate the analytical pump, if required. Refer to the pump manual for
instructions.
Detector Calibration
Calibrate the detector, if required. Refer to the detector manual for
instructions.
7.2.6
Set Up the Dionex Integral Analyzers in Chromeleon 7 PA
In Chromeleon 7:
•
Create instrument methods and processing methods for each Dionex
Integral
system.
•
Run a standard, and then edit component information for the method.
•
Calibrate the method for the first time. (The method can subsequently
be calibrated in Chromeleon 7 PA.)
2
2
1
1
C
V
C
V
2
1
1
2
V
C
V
C
2000
mL
50
L
25
V
V
D
2
1
f
μ
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...