Dionex Integral Systems Operator’s Manual
46
Doc. 065250-06 03/2020
5. Plug the analog input cable (P/N 070171) into an analog input connector on
the main SP electronics board.
NOTE Although Analog Input 4 is usually reserved for
monitoring of the stir bar in the dilution vessel mixer
(see
), it can be used for a different function
by simply changing the position of a jumper on the main
SP electronics board. For instructions, see
the position of the Analog Input 4 jumper:” on page 47
6. Attach wires to the 2-position connector plug on the other end of the cable.
To attach a wire to the plug: Strip the end of the wire, insert it into the plug,
and use a screwdriver to tighten the locking screw.
7. Route the cable to the other device, and then connect the free ends of the wires
to the appropriate connector pins on the device.
8. Close and secure the SP service door.
9. If you removed the SP from the AE rear panel:
a.
Lift up the SP, align the four hooks on the rear of the SP with the
mounting rails on the AE rear panel, and hang the SP on the rails.
b. Replace any modules that were removed from the lower shelf of the AE.
c.
Replace the two access panels on the exterior of the AE.
10. Plug in the power cord and turn on the power.
11. Open the Chromeleon 7 Instrument Configuration Manager and open the SP
Properties
dialog box. The check box for the connected analog input is
enabled on the
Analog Inputs
page. The name assigned to the input
corresponds to the number of the connector on the main SP electronics board.
12. Save the configuration.
13. Calibrate the analog input (see
When attaching wires to the connector plug, be careful not to allow
stray strands of wire to short to an adjoining position on the
connector.
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...