Assign TTL Input Control Types (Optional)
If you connected a TTL input, verify that the correct input action and
control type are selected. Select different settings, if necessary.
The SP TTL inputs respond to four types of signals. The default control
type,
Normal Edge
, is compatible with the output signals provided by
Dionex modules.
If the device connected to the SP does not send a normal edge signal,
follow the steps below to select the appropriate control type. To determine
the correct type, refer to the documentation provided with the controlling
device and to the information in
“TTL Input Control Types” on page 167
.
To select an input control type:
1. Open the Chromeleon 7 Instrument Configuration Manager.
2. Double-click the SP icon under the instrument name. The
Properties
dialog box appears.
3. Click the
TTL Inputs
tab.
4. Select the name of the input and press the
F2
key. The
Device
Configuration
dialog box for the TTL input appears.
5. Select the
Mode
and click
OK
.
TTL Input Control Types
•
Normal Edge
: In normal edge
operation, the negative (falling)
edge of a signal turns on the
action.
The function of the positive
(rising) edge depends on the action: For actions that have two options
(for example, on/off), the rising edge turns off the action. However,
for functions with only one option, the rising edge has no effect.
For example, for the
Bypass Streams
,
Panel Light
, and
Relay Out
actions, the action remains on as long as the TTL input is low. If the
input returns to high, the action is turned off. For example, if an
external device triggers a panel light, the light remains on as long as
the input is at 0 V (open). When the input returns to +5 V (closed), the
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...