Doc. 065250-06 03/2020
61
4 • SS Stream Selector Description
4.1
SS Overview
The SS Stream Selector (see
) allows selection of a sample from
multiple sample streams. The SS can be configured with up to three stream
selection valves, with each valve connected to up to seven sample streams. This
allows sampling from one of 7, 14, or 21 streams, depending on the number of
valves installed in the SS.
4.2
Stream Selection
SS stream selection valves are 17-port, 8-position valves in either PEEK
(P/N 068535) or stainless steel (P/N 068540). Sample stream inlet lines enter the
SS through fittings on the side of the SS and connect to inlet ports on the stream
Figure 4-1. SS Stream Selector with One Stream Selection Valve
Valve A
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...