10 • Service
Doc. 065250-06 03/2020
141
4. The fuse drawer is located above the power
receptacle (see
locks the fuse drawer in place. Using a small
screwdriver, press the tab
in
and
then up
to
release the fuse drawer.
5. Pull the fuse drawer out of the rear panel
and remove the old fuses.
6. Replace the two fuses with new fast-blow
IEC 127 fuses rated 3.15 A (P/N 954745).
Thermo Fisher Scientific recommends
always replacing
both
fuses.
Figure 10-13. Fuse Drawer
7. Insert the fuse drawer into the rear panel and press until the drawer snaps into
place.
8. Reconnect the main power cord and turn on the power.
Fuse
Drawer
Tab