9 • Troubleshooting
Doc. 065250-06 03/2020
105
If a sample preparation command is assigned a specific retention time,
position the command outside the sample preparation command block.
This error occurs if the instrument method includes a
DelaySP
command, but
the command is not inside the sample preparation command block. See
for more information.
To troubleshoot:
Move the command inside the sample preparation command block.
–or–
Delete the sample preparation command from the method.
This error may indicate a problem in the dilution pump electronics.
To troubleshoot:
1. Turn off the pump flow.
2. Turn off the SP power briefly by pressing the SP power switch.
3. Turn on the SP power again and restart the pump flow.
4. If the problem persists, contact Technical Support for Dionex products for
assistance.
DelaySP command must occur within BeginSPSamplePrep/
EndSPSamplePrep command block
Dilution pump driver is over current
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...