Doc. 065250-06 03/2020
143
A • Specifications
A.1
Environmental (All Analyzer Components)
A.2
AE Analyzer Enclosure and LE Liquids Enclosure
A.2.1
Electrical
Operating
Temperature
10 to 40 °C (50 to 104 °F)
Humidity
5% to 95% relative humidity, noncondensing
AE Main Power
100 to 240 Vac, 50/60 Hz (Auto-sensing power supply; no manual
voltage or frequency adjustment required)
Typical input power: 10 W (AE only, without other devices
connected)
Line draw: Actual current draw depends on the devices installed in
the enclosure; total current not to exceed 20 A at 120 Vac
AE Circuit
Breaker
20 A
Short Circuit
Current Rating
5 kA