Dionex Integral Systems Operator’s Manual
92
Doc. 065250-06 03/2020
Notes
•
Commands for the SP and SS must be added in the Script Editor view
in the Instrument Method Editor. Commands for other Dionex
Integral system components can be added in the Instrument Method
Wizard.
•
To ensure that sample preparation works correctly, method
commands must be grouped into one block of commands for the SP
and another for the SS. The
BeginSPSamplePrep
command must be
the first command in the block, while the
EndSPSamplePrep
command must be the last command. For more information, refer to
the Chromeleon 7 Help.
In Chromeleon 7 PA:
•
Configure the Dionex Integral analyzers, instruments, and streams.
Refer to the Chromeleon 7 PA Help for detailed instructions.
7.3
Routine Startup and Operation
Routine operation consists of first confirming that all hardware is operating
properly, and then running instrument methods and sequences to control sample
analysis. Refer to the Chromeleon 7 PA Help for detailed operating instructions.
7.3.1
Routine Startup
1. Confirm that the power is on for all system components.
2. Confirm that the computer is on.
3. Check that the Chromeleon 7 Server is running.
4. Confirm that all water and gas utilities are on and adjusted to their
proper pressures.
5. Confirm that all mobile phases, standards, and reagents are supplied.
6. Start Chromeleon 7 PA.
7. Load an appropriate sequence.
8. Verify the following:
•
The mobile phase flow rates are correct.
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...