Dionex Integral Systems Operator’s Manual
64
Doc. 065250-06 03/2020
4.4
SS Power and Computer Connections
Receptacles for connecting a power cord and USB cable are on the side of the SS
(see
Fuse Holder and Power Receptacle
The fuse cartridge contains two fast-blow IEC 127 fuses rated 3.15 A
(P/N 954745). For instructions on how to change the fuses, see
The power cord plugs into the IEC 320 three-prong receptacle. Connect the power
cord from this connector to a grounded power source.
•
If the SS is mounted on an AE or installed inside an AE, connect the power
cord to the power outlet strip inside the AE (see
). The AE power
outlet strip is a grounded receptacle.
•
When the SS is mounted on the side of the AE or installed inside the AE, the
AE power cord is the main disconnect device. In addition, the AE main power
switch (see
Emergency Off
used to control the power to the SS.
Figure 4-3. SS USB and Power Receptacles
SHOCK HAZARD—If a grounded receptacle is not used, a shock
hazard may result. Do not operate or connect to AC power mains
without earthed ground connections.
The power supply cord is used as the main disconnect device. Make
sure the socket-outlet is located near the module and is easily
accessible.
Fuse Holder
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...