Dionex Integral Systems Operator’s Manual
134
Doc. 065250-06 03/2020
10.7 Accessing the Main SP Electronics Board
Refer to the appropriate instructions below when performing a service procedure
that requires access to the main SP electronics board. The instructions vary
slightly, depending on the type of SP enclosure.
If the SP is an external enclosure:
1. Turn off the SP power.
2. Disconnect the power cord.
3. Unlatch the two handles on the service door of the SP and open the door. The
main SP electronics board is mounted on the inside of the door (see
Figure 10-7. SP External Enclosure: Service Door Open