3 • SP Sample Preparer Description
Doc. 065250-06 03/2020
51
3.10 Flow Sensor (Option)
The optional Flow Sensor Kit (P/N 068564) provides a sensor for measuring the
liquid flow rate in a stream. Flow rate readings from the sensor are displayed on
the SP ePanel in Chromeleon 7. The sensor can measure flow from 0.2 to
5.0 mL/min. The operating pressure for the flow sensor must be less than
0.28 MPa (40 psi). The resolution of data is 16 bits and the accuracy is 5% of the
measured value. Up to two flow sensors can be installed in the SP.
High and low limits for the flow rate are entered in the Chromeleon 7 Instrument
Configuration Manager. If the flow rate is outside the specified range, the sensor
reports the error to Chromeleon 7 and an error message is displayed in the audit
trail.
To install the flow sensor:
1. Turn off the SP power.
2. Disconnect the power cord.
3. Using the screws and washers provided in the Flow Sensor Kit, attach the
flow sensor to the mounting plate as shown in
. Make sure the
nylon washers are installed between the board on the flow sensor and the
mounting plate.
Figure 3-31. Flow Sensor
Flow Sensor
Mounting Plate
Mounting
Panel
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...