Dionex Integral Systems Operator’s Manual
186
Doc. 065250-06 03/2020
4. Close and secure the SP service door.
E.2.3
Assembling the Peristaltic Tubing and Fittings
1. Install the peristaltic tubing adapter parts on the PharMed tubing in
the order and orientation shown in
. Use the retainer sleeve
with an inside diameter that is closest in size to the outside diameter
of the tubing.
2. Push the point of the adapter body into the tubing until the tubing is
flush with the adapter body (see
3. Slide the cap over the retainer sleeve and tighten the cap onto the
body.
4. Repeat
through
to install the tubing adapter on the other
end of the tubing.
5. Lift open the door of the peristaltic pump.
Figure E-4. Peristaltic Tubing Adapter Connections
Retainer Sleeve
Adapter Body
Cap
Peristaltic Tubing
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...