29
Operation Procedure
Operation indicator
Checks
• The message “Close the door after
setting the tube” and guidance displays
on the LCD.
• The AC icon
is displayed.
• The battery icon is displayed, not
indicating the sub-battery failure.
• The operation indicator is off.
Checking an Infusion Set
Infusion set
Checks
• Only the dedicated infusion set should be used. (See page 19.)
• Check set for damage.
Preparing an Infusion Set and Priming
Pump
Manual roller
clamp
1)
Place the manual roller clamp of the infusion set on the
lower side of the infusion pump, and close the manual roller
clamp.
Note
• Setting up the manual roller clamp on the lower side of the infusion pump allows
suppression of bolus infusion (temporary excessive infusion of drug solution) at the
time that any cause of occlusion is removed. In addition, when an Air-in-line alarm
is issued, air bubbles can be released into the drip chamber by closing the manual
roller clamp.
Drip
chamber
Drug
solution
Approx. 1/3
2)
Insert the spike of the infusion set into the drug solution
container and squeeze the drip chamber so that the drug
solution fills it to approx. 1/3 of the drip chamber.
Open
3)
Open the manual roller clamp to allow the drug solution to
fill the line. (Priming)