22
Preparation
Pole clamp
●
Base portion of the pole clamp
●
Pole clamp knob
●
Pole clamp lock portion
●
Clamp portion of
the pole clamp
1. There is no damage to any part of the pole clamp and that the clamp is clean.
2. When attaching pole clamp to the pump, it should lock in place firmly with an audible click.
3. Clean periodically even if there is no adhered drug solution.
2)
The following self-check (self-diagnosis) operations should be performed when the power is
turned on using the internal battery while the door is left open without an infusion set installed.
1. The operation indicator flashes and the fingers move slightly.
2. The message “Self-check” is displayed on the LCD, and the LCD flashes.
3. The buzzer sounds.
4. Infusion sets guidance is displayed, and when the door is closed, the icons “Air-in-line detector
”, “Upper occlusion
detector
” and “Lower occlusion detector
” flash on the LCD.
3)
Neither Battery alarm or Power Failure alarm has been issued.
1. The level of battery icon on the LCD has two bars (green)
or more.
2. Neither Battery alarm (yellow) or Power Failure alarm (red) is displayed on the LCD.
4)
When the AC power cable is connected, the LCD displays AC icon
.
1. When the AC power cable is inserted into the AC inlet on the back of the main unit completely, the AC icon
is
displayed.
2. If the internal battery is not fully charged, fully charge it.
Note
• During charging,
and
flash alternately.
5)
When the AC power cable is connected, there is no looseness or any fault.
• The AC power cable is completely inserted into the AC inlet of the main unit.