Device Characteristics
Occlusion Characteristics
Typical occlusion detection ability data are occlusion pressure, Occlusion alarm occurrence time, and post-occlusion removal
bolus volume.
The data below is a representative example measured for flow rates of 1 mL/h and 25 mL/h, with a three-stage occlusion pressure
setting value (1, 5, 10), a new TERUFUSION Solution Administration Set for Infusion pump, and the occlusion clamp position at
one meter below the pump. The measured values may vary depending on the test conditions.
Flow rate
(mL/h)
Occlusion
pressure
setting value
Occlusion pressure
Time to activate an
Occlusion alarm time
(min)
Bolus volume
(mL)
(kPa)
(kgf/cm
2
)
1
1
26 0.27
30.6
0.21
5
64 0.65
71.2
0.47
10
106 1.08
127.2
1.06
25
1
29
0.30
1.1
0.33
5
67
0.68
2.5
0.54
10
109
1.11
4.4
1.11