39
Operation Procedure
Starting Solution Delivery
Start switch
Operation indicator (green flashing)
•
Check the flow rate setting and press the Start switch to
start solution delivery.
Notes
• The buzzer sounds.
• The operation indicator flashes green.
• Dripping inside the drip chamber starts.
• The volume delivered increases and the remaining time decreases as shown
below.
• Flow rate can be changed during solution delivery. Rotate the Dial in the flow rate
display to change the flow rate and the confirmation message appears. Press the
Start switch to start solution delivery at the changed flow rate.
Volume delivered
Remaining time
Notes
• When the pump is fully set but not started, after 2 minutes the LCD displays the
Start Reminder screen (yellow) and the buzzer sounds.
• Use the standby function (see page 47) if you want to temporarily extend the
Start Reminder.
When Infusion Cannot Start
The alarm sounds and the following warning display. (See page 91.)
No Flow Rate alarm (yellow)
The flow rate has not been set. Set the flow rate and press the Start switch again.
No VTBI alarm (yellow)
The VTBI has not been set. Set the VTBI and press the Start switch again.