Prior to the initial use of the pump, set the date and time to record history correctly. (see page73.)
Before use, carefully read “Precautions” and the labels attached to the product.
Prior to the Pump being Used
Connect to the AC power supply and charge (8 hours or more) with the power turned off.
Prior-to-use Inspection
Check the following points before use.
1)
There is no damage to the main unit, AC power cable, pole clamp or drip sensor, and that there
is no malfunction due to adhered drug solution.
Main unit/AC power cable/Drip sensor
1
2
4
3
2
2
1. There is no damage, dent or adhered drug solution on the AC inlet and AC power cable.
2. There is no damage to the drip sensor, the drip sensor connector cover, the plug of drip sensor, Lower tube holder/Upper
tube holder and RS-232C connector cover (if RS-232C is connected.) (TE-LF632 only).
3. There is no damage to the handle, the tube guides, the fingers, the buffer plate, the various detectors and the tube
clamp.
4. There is no adhered drug solution on the Dial, it turns smoothly, and flow rate setting can be set when the power is on.
If there is adhered drug solution, promptly clean in accordance with the directions on page 86.
Preparation