38
Operation Procedure
Opening the Manual Roller Clamp on the Infusion Set
Fully open
Solution must not keep dripping.
•
Fully open the manual roller clamp of the infusion set.
Checks
• No drug solution drips from the needle tip or the inside of the drip chamber of the
infusion set.
• The flow rate setting is correct.
Cautions
• After the infusion set is attached, the door is closed and the manual roller
clamp is opened, if any dripping of drug solution is observed, check the
infusion set (the dedicated infusion set), the attachment status of the tube
(correctly attached), and any fault in the infusion set and devices (whether or
not damaged).
• If none of the defects are identified and dripping is observed, it is likely
that the pump is defective. Immediately stop using the product and contact
TERUMO trained service technicians.
Inserting a Needle
Vein
1)
Check
• The pump is stopped.
2)
Insert the needle according to the internal hospital
procedures.
Warning
• This product does not have an alarm function to detect extravascular infusion
due to intravenous needle coming out of vein. Periodically check the puncture
site.