Troubleshooting
Nearly Empty alarm
Occurrence
Cause
Action
• LCD: The Nearly Empty alarm
display lights up yellow.
• Operation indicator:
Flashes green (during solution
delivery) or off (stopped)
• Buzzer: Activates.
• Pump: No changes to the
status.
• The drug solution is running
out.
• To continue solution delivery, replace the drug solution
container with a new one.
Re-alarm
Occurrence
Cause
Action
• LCD: The respective alarm
displays flash/light up (the
flashing/lighting up colours
are the same as those at the
respective alarms).
• Operation indicator: Same as
for when the alarms are first
activated.
• Buzzer: Activates.
• Pump: No change to the
status.
• Operation was not started
within 2 minutes after the
alarm was muted.
• Follow the corrective action of each alarm.
Other problems
Maintenance timer
Occurrence
Cause
Action
• The Maintenance timer
indicator
lights up.
• It is time to perform
maintenance.
• Contact TERUMO trained service technicians.
AC power detector function
Occurrence
Cause
Action
• The pump cannot be turned on
when using AC power.
• The AC icon
is not displayed
even after the AC power cable
is connected.
• Check the connection of the AC power cable to the
pump and the power supply. (See page 25.)
• The AC power cable is
damaged.
• The AC power cable is
internally disconnected.
• Replace the AC power cable.