106
Main unit
Product name
TERUFUSION Infusion Pump Type LF3
Model
TE-LF630
Catalogue number
TE
*
LF630Nxx, TE
*
LF632Nxx
Alphanumeric characters are entered for x in the catalogue number.
(See page 110.)
Compatible infusion sets
Sterilized specified infusion sets (See page 19.)
Brand of infusion set
TERUMO
Flow rate setting range
• Setting range
3.00 to 300.00 mL/h (when setting is 20 drop/mL)
1.00 to 100.00 mL/h (when setting is 60 drop/mL)
• Step
1.00 mL/h step*
*:
Step can be used when flow rate is set by the Dial.
(It will be 0.00 mL/h when the power is turned on.)
VTBI setting range
0.10 to 100.00 mL (0.10 mL step)
100.00 to 9999.00 mL (1.00 mL step)
“----.--” (Unlimited)
(It will be 0.00 mL when the power is turned on.)
VTBI time setting range
1 min to 99 hours 59 min (1 min step)
“-- --” (Unlimited)
Hands On Bolus flow rate setting range
100.00 to 300.00 mL/h (when setting is 20 drop/mL)
100.00 mL/h (when setting is 60 drop/mL)
(100.00 mL/h step)
Hands Free Bolus dose rate setting range
0.01 to 10.00 mL (0.01 mL step)
10.00 to 100.00 mL (0.10 mL step)
100.00 to 999.00 mL (1.00 mL step)
Hands Free Bolus dose time setting range
1 s to 60 min 00 s (1 s step)
Volume delivered display range
0.00 to
10.00 mL (0.01 mL step)
10.00 to 100.00 mL (0.10 mL step)
100.00 to 9999.00 mL (1.00 mL step)
(It will be 0.00 mL when the power is turned on.)
Flow rate accuracy
Within ±10%
(The hourly precision after one hour has passed since the start of solution
delivery, in the case where water or a physiological saline solution is used
at a constant flow rate of minimum flow rate or more (ambient temperature:
23±2°C) according to IEC 60601-2-24:2012 (EN 60601-2-24:2015) standard.)
Occlusion detection pressure
• Upper occlusion detection pressure
-100 to -30 kPa
• Lower occlusion detection pressure
30 to 120 kPa(set value)
The lower occlusion detection pressure can be set in 10 levels.
Purge flow rate
Approx.300 mL/h or more (depending on the infusion set to be used)
Alarms
Upper Occlusion alarm, Lower Occlusion alarm, Nearly Empty alarm, Air-in-line
alarm, Door alarm, Flow Rate Abnormality alarm, Free Flow alarm, Line Empty
alarm, Battery alarm, Shutdown Notice alarm, Power Failure alarm, Re-alarm,
Start Reminder, No Flow Rate alarm, No VTBI alarm, Flow Rate/VTBI Volume
Judgment alarm, Completion alarm
Safety functions
Occlusion detection
pressure select
function
Selects the lower occlusion detection pressure setting values in 10 levels.
Air-in-line alarm
sensitivity select
function
Changes the detection sensitivity of the air-in-line detector.
Specifications