103
Troubleshooting
Battery Replacement Time/Failure alarm
Occurrence
Cause
Action
• The battery operation time is
short even after fully charged.
• The battery icon
is
displayed.
• The battery icon does not
display 5 bars after charging for
8 hours or more.
• The battery has deteriorated
or is damaged.
• Replace the internal battery.
•
and the battery charge
level appear alternately after
more than 8 hours of charging.
• The sub-battery has
deteriorated or is damaged.
• Replace the sub-battery.
Flow rate
Occurrence
Cause
Action
• Pressing the Start switch does
not start solution delivery.
• The infusion line is not open.
• The infusion set is not installed
correctly.
• Re-set the infusion set. (See page 30.)
• The infusion set is occluded.
• Eliminate the cause of occlusion.
(See pages 94 and 95.)
• Flow rate accuracy is low.
• The infusion set is not installed
correctly.
• Check the placement of the infusion set in the pump.
(See page 30.)
• Drug solution with high-
viscosity (e.g. glucose
injection) or low-viscosity is
used.
• Considering the size of drops, correct the flow rate and
start solution delivery. (See page 44.)
• The tube has been attached
to the same place for a long
period.
• Check that the tube, when pressed with fingers then
released, recovers to the normal state. If not, adjust
placement or replace the infusion set.
• An infusion set other than
those specified is being used.
• If an infusion set other than the ones specified is
being used, replace it with the specified infusion set.
(See page 19.)
• The flow rate is not set
correctly.
• Correct the flow rate setting.
• There is large height
difference between the pump
and patient.
• Adjust the patient to the pump height differential.
The door will not close
Occurrence
Cause
Action
• Even when you try to shut the
door, you cannot lock it.
• The tube is not installed
properly.
• Re-set the infusion set. (See page 30.)
External communication
Occurrence
Cause
Action
• Cannot communicate
(RS-232C)
• The baud rate, stop bit, parity,
or device ID number may not
be appropriate.
• Check the settings of the connected device.
(See page 75.)
• Cables are not securely
connected.
• Reconnect the cable. (See page 75.)