After Use
How to Clean Components
After use, clean all components as described below.
Wipe dirt off gently with a cotton swab or
damp cloth (not wet).
Lightly wipe off dirt with a cotton swab or damp
cloth (not wet).
Wipe dirt off gently with a cotton swab or damp cloth
(not wet).
* Do not use ethanol to disinfect the Air-in-line detector
and the occlusion detector.
Cleaning the panel surface
Wipe with a damp (not wet) soft cloth.
Cleaning the outside of the infusion pump
Cleaning the Dial
Cleaning the buffer plate
Cleaning the tube clamp
Cleaning the air-in-line detector and the occlusion detectors
The Dial is fixed to the body of the infusion
pump with a magnet. Remove it following
the procedure below before cleaning.
How to remove the Dial
1. Turn the groove of the Dial to the rear of
the device.
2. Place a flathead screwdriver in the
groove of the Dial and push it outwards.
3. Remove the Dial and wipe clean the
body, and the Dial.
Grease is applied to the shaft. Do not
wipe it off.
4. When you have finished cleaning, fit the
dial back onto the body.
5. After fitting, turn on the power.
Check that setting for flow rate etc. can
be performed.
Wipe with a soft damp cloth. When
disinfecting, use a designated disinfectant.
Cautions
• Do not rub the panel surface with a hard object.
• Pay attention not to damage the tube attachment surface when cleaning.
• Do not rub with a hard object such as forceps or with a sharp object etc.