31
Operation Procedure
Closing the Door
Door lock lever
Door
Correctly set
Wrongly set
Correctly set
Wrongly set
1)
When closing the door, lock the door with the door lock
lever.
Check
• The message “Door Closed 20 drop/ mL” appears on the LCD and then the screen
changes to the brand of infusion set selecting display.
(Only when the drip volume is set to 20 drop/mL)
Brand of infusion set tab
• The tube is not caught in the door.
• All the sensor icons are turned off.
Caution
• This product has a safety system making it hard to close the door when the
tube has not been set straight (AIS (Anti Irregular Set) system). Forcing the
door closed may result in damage to the device and/or tube.
If the door is hard to close, check that the tube is set correctly instead of
forcing the door closed.
Lower tube holder
Upper tube holder
2)
Set the tube to the Upper and Lower tube holders.
Caution
• When setting the tube, do not pull excessively hard. The tube may be
deformed, and therefore the original functionality or performance of this
product may not be achieved. (flow rate error and various alarm functions etc.)