82
About External Communication Functions
Item
Size (Byte)
Details
Operation monitor
1
b0-b1 (low order)
Lower Occlusion monitor (0x0: 0%, 0x1: 25%, 0x2:
50%, 0x3: 75%)
b2
(Reserved)
b3
Alarm sound status (0: Without sound, 1: With
sound)
b4
Keypad lock status (0: Not locked, 1: Locked)
b5
Monitor for the door (0: Close, 1: Open)
b6
Drip sensor status (0: Connected, 1: Not
connected)
b7 (high order)
(Reserved)
One bit used for b2 - b7. (“1” indicates active.)
Infusion set status
1
0x0
(Reserved)
0x1
20 drops
0x2
60 drops
0x3
(Reserved)
0x4
(Reserved)
0x5
(Reserved)
0x6
(Reserved)
0x7-0xFF
(Reserved)
Flow rate
3
Value of 100 times the set flow rate
Volume delivered
3
Value of 100 times the volume delivered
VTBI
3
Value of 10 times the VTBI
Infinite: 0xFFFFFF
(Reserved)
3
(Reserved)
(Reserved)
1
(Reserved)
(Reserved)
2
(Reserved)
(Reserved)
3
(Reserved)
(Reserved)
1
(Reserved)
(Reserved)
2
(Reserved)
(Reserved)
2
(Reserved)
Occlusion pressure setting 1
bit0-bit3: Setting of occlusion pressure. 1 to 10.
Occlusion pressure range: 0x01-0x0A
bit7: Setting of occlusion detection pressure 1 = In 10 levels
(bit4-bit6: Reserved)
Hands On Bolus flow rate
2
100.00 to 300.00 mL/h (100 mL/h step)
Hands Free Bolus VTBI
3
Value of 100 times the bolus VTBI
Invalid value (when the function is disabled): 0x000000
Hands Free Bolus dose
time
2
0 to 3600 s (1 s step)
Invalid value (when the function is disabled): 0x0000
Hands Free Bolus flow
rate
3
Value of 100 times the flow rate
Invalid value (when the function is disabled): 0x000000
Infusion set brand number 1
Brand number of the infusion set selected
Total number of bytes
46
* For more details, contact TERUMO trained service technicians.