36
Operation Procedure
Setting VTBI
Display
select switch
1)
Press the Display select switch to select the VTBI display.
Notes
• The VTBI tab
is selected.
• The flow rate display shows the VTBI display and the VTBI time display.
• The VTBI display appears in dark blue (selected).
VTBI time display
VTBI display
Flow rate tab
VTBI tab
Dial
Increases value
Decreases
value
2)
Use the Dial to adjust the VTBI.
Note
• When the flow rate has already been entered, the VTBI time display indicates the
remaining time and VTBI time calculated from the flow rate and VTBI.
Remaining time
Start up icon
VTBI time
Notes
• The VTBI can be set from 0.10 mL to 9999.00 mL and “----.--” (Unlimited).
• If not operated for 10 seconds, the input area moves to the flow rate display.
• VTBI can be changed only when delivery is stopped.
• To set to unlimited, rotate half in the direction of decreasing the Dial when VTBI is “0”.
• Rotating the Dial while pressing the Stop switch allows changing the setting at the rate of 100 times.
Caution
• In the case of a solution delivery with no setting for the VTBI (“----.--”: Unlimited), the pump will not stop until an alarm is
issued. The pump should be stopped before running out of the drug solution. It is recommended to set a VTBI slightly less
than the volume of the drug solution container.