Changing Settings on the Menu
Dial
Increases value
Decreases
value
9)
Turn the Dial to set the Bolus VTBI time or the Bolus flow
rate.
Notes
• The Bolus VTBI time or the Bolus flow rate (mL/h) has been calculated from the
Bolus VTBI, and is displayed.
• The ready for bolus icon is displayed.
Ready for bolus icon
Bolus flow rate
Bolus VTBI time
• The Bolus flow rate can be set within the range of the following values.
20 drop/mL: 3.00 to 300.00 mL/h
60 drop/mL: 1.00 to 100.00 mL/h
• The Bolus VTBI time can be set within the range below.
Bolus VTBI time: 00min01s to 60min00s.
Caution
• When using the Hands Free Bolus function, the setting cannot take a flow rate calculation result outside the flow rate
setting range, and therefore it should be reconfigured. [----.--] or [0.00] is displayed, not allowing solution delivery.
Operation indicator (green flashing)
Purge switch
10) Press the Purge switch to start the bolus.
Notes
• The operation indicator flashes green.
• The bolus volume delivered increases.
• The alarm melody sounds during solution delivery.
Bolus volume delivered