45
Other Operation Procedures
Purging
When the infusion is stopped, the solution delivery can be purged.
Purge switch
Operation indicator (green flashing)
•
Press and hold the Purge switch when the solution delivery
is stopped.
Notes
• The buzzer sounds and the operation indicator flashes green.
• The flow rate is switched to purge flow rate.
• If any other switch is pressed while the Purge switch has been pressed, the safe
function detects it as an abnormal operation and stops purging.
Flow rate
Notes
• The volume purged is not displayed on LCD.
• The upper limit of purge volume can be set. Contact TERUMO trained service technicians.
• The solution delivery cannot be purged when an alarm is not cleared.
Cautions
• When purging the solution, note that the volume purged is not added to the volume delivered.
• Purging is intended for priming the infusion line without patient being connected.