47
Other Operation Procedures
Standby Function
The standby function can extend the Start Reminder.
Note
• The infusion set is attached and the solution delivery is ready to be started.
Back/Mute
switch
2 seconds
or more
•
Press and hold the Back/Mute switch for 2 seconds or more.
(The standby mode is entered.)
Note
• The LCD displays the message “Stand By”.
Notes
• The default standby time is 24 hours. To change this, contact TERUMO trained service technicians (the time can be changed to
24 hours or less).
• Turning the Dial or pressing any switch cancels standby mode.
• The standby mode cannot be entered when an alarm has been issued. Eliminate the cause of the alarm first.
Setting Special Functions
The following settings are selectable by changing the internal settings of this product. Taking account of a subsequent impact on
product performance, contact TERUMO trained service technicians.
Setting item
Description of function
Setting dial
acceleration
function
Rotating the Dial while pressing the Stop switch accelerates the setting rate to set items such as the flow
rate. This function is not available during solution delivery. Pressing the Stop switch stops the solution
delivery.
• Flow rate and VTBI: change the rate to 100-step increments
• Date and time: change the rate to hour and month
• History: change the rate to page feeding
Infusion set brand
setting
The infusion brand and drip volume are selected after the power is turned on and the set guidance has
finished. This function is ON by default setting.
Flow rate range
setting function A
(Soft limits)
Flow rate range
setting function B
(Hard limits)
Sets the limitation of the flow rate in 2 levels; the soft limit (can be exceeded after confirmation) and the
hard limit (cannot be exceeded).
• When setting the flow rate over the soft limit, the flow rate will be displayed in orange.
Change the KVO
flow rate
Allows the KVO flow rate to be set within the following ranges
• When using an infusion set of 20 drop/mL: 3.00 mL/h (Fixed)
• When using an infusion set of 60 drop/mL: 1.00 to 3.00 mL/h
Hands Free Bolus
setting
When bolus volume and bolus time are set, flow rate is calculated and the bolus infusion is available. (For
details, see page 61.)
Alarm melody
select function
Changes the alarm melody from Pattern 1 to Pattern 2.
The default setting is Pattern 1.
Night mode setting If the device is left non-operated for 30 seconds between 21:00 to 5:00 (customizable), the brightness can
be automatically lowered by two levels (customizable).