109
Specifications
Power supply
AC100 to 240V, 50 to 60Hz
Battery: Lithium-ion battery
• Continuous use period: Approx. 5 hours (applies when solution is delivered
continuously at 25 mL/h in ambient temperature of
25°C with new fully charged battery)
• Charging time:
≥
8 hours (when charged with AC power supply with
the power turned off)
The battery can be charged to approx. 80% in
3 hours.
Sub-battery: Nickel-metal-hydride battery (for a Power Failure alarm in case of
both power supplies are lost)
• Charging time:
≥
8 hours (when charged with a new battery and AC power
supply in ambient temperature of 25°C when the power is off)
Power consumption
28 VA
Classification
Electric shock
protection
Class I equipment and internally powered equipment, Defibrillation-proof type
CF applied part
Mode of operation
Continuous operation
Protection against
harmful ingress of
water or particulate
matter
IP22 (Vertically dripping water shall have no harmful effect when the enclosure
is tilted at an angle up to 15° from its normal position.)
Dimensions
Without external communication (RS-232C):
253 mm (W)
*
1
x 120 mm (H)
*
1
x 102 mm (D)
*
2
With external communication (RS-232C):
253 mm (W)
*
1
x 120 mm (H)
*
1
x 121 mm (D)
*
2
*
1
: Excluding protrusions
*
2
: Excluding protrusions, pole clamp and moving range
Unit weight
Approx. 2.0 kg
Standard accessories
Drip sensor, AC power cable, pole clamp (Screw type), instruction manual
• This product is in compliance with EMC (electromagnetic compatibility) standard IEC 60601-1-2:2007 (EN 60601-1-2:2007)
(CISPR group classification and class classification are Group 1 and Class B). It is also in compliance with the EMC level
required by IEC 60601-2-24:2012 (EN 60601-2-24:2015).
• Conformity standard and directive
IEC 60601-1:2005+A1:2012 (EN 60601-1:2006+A1:2013)
IEC 60601-1-2:2007 (EN 60601-1-2:2007)
IEC 60601-1-6:2010+A1:2013 (EN 60601-1-6:2010+A1:2015)
IEC 60601-1-8:2006+A1:2012 (EN 60601-1-8:2007+A1:2013)
IEC 60601-2-24:2012 (EN 60601-2-24:2015)
MDD (Medical Device Directive) 93/42/EEC (Class IIb)
Note
• If an AC power cable is not included with the pump unit, please contact TERUMO trained service technicians to receive the cable
suited for your region.