50
Changing Settings on the Menu
Occlusion Detection Pressure
The occlusion detection pressure can be set in 10 levels.
Setting
Occlusion detection pressure
Setting
Occlusion detection pressure
1
± 30 kPa
6
± 80 kPa
2
± 40 kPa
7
± 90 kPa
3
± 50 kPa
8
± 100 kPa
4
± 60 kPa
9
± 110 kPa
5
± 70 kPa (Default)
10
± 120 kPa
Warning
• After the occlusion detection pressure is selected, check that the setting has been adjusted correctly.
Notes
• If using on “Level 10”, the pressure of an infusion line increases at occlusion, possibly resulting in the infusion line being
disconnected. Periodically check that the connection of the infusion line is secured.
• If using on “Level 1”, an Occlusion alarm is frequently issued, especially when using a drug solution with high viscosity, possibly
resulting in infusion being interrupted. In such cases, review the set value of the occlusion pressure by using the graph of the
history of occlusion pressure value as a guide.
• Although the Occlusion alarm sounds, the internal pressure level may not reach the level of 75% or higher (the Occlusion icon in
orange) due to the activated bolus reduction function.
• Use the occlusion icon
as follows:
(1) To check the in-line pressure when using a small diameter administration set and/or infusing a highly viscous solution.
(2) To make sure the infusion line has been opened.
2 seconds
or more
Display
select switch
1)
Press and hold the Display select switch for 2 seconds or
more.
Note
• The menu 1 screen is displayed and the icons are ready to be selected.
Occlusion detection pressure tab
Confirmation
switch
2)
Press the Confirmation switch.
Note
• The existing setting (light blue) changes to the selected status (dark blue).