48
Other Operation Procedures
Setting item
Description of function
Change the
standby duration
Up to 24 hours with 10 minute intervals.
Air-in-line Alarm
(Accumulated
bubble detection
function)
Increase the sensitivity level for bubbles accumulated. The alarm is issued if accumulated microbubbles
exceed 1 mL within 15 minutes.
* Air-in-line alarm may occur more frequently when you increase the accumulated bubble detection.
Purge volume
upper limit setting
function
The upper limit of purge volume can be set. The purging stops when the set upper limit is reached.
Bolus volume
upper limit setting
function
When the Hands On Bolus function is enabled, the upper limit of bolus volume can be set. Bolus stops
when the set upper limit is reached.
Set value memory
function
Saves the current set values (e.g. flow rate, VTBI) before the power is turned off.
Remaining volume
detection setting
function
Selects the time to activate the Nearly Empty alarm from Disabled (no alarm), 5 min., 10 min., and 15 min.